Defibrillator
Output: Biphasic Truncated exponential
Energy Sequence: 150, 150, 200J
Charge Time: 4sec. to150J
8 sec. to 200J
Analysis Time: 4-16 sec
Combined Analysis and Charge Time from power up after 6 shocks
From start of analysis to shock ready:
Less than 20 seconds to 150J
Less than 30 seconds to 200J
Audible Prompts: 20 audible prompts
Visual Prompts: LED prompts
Controls: Two buttons – On/Off, Shock
Output Energy Accuracy:
±15% or ±3J (whichever is greater) to any impedance from 25 to 175Ω
Battery
Non-Rechargeable LiMnO2 12V, 3.0 Ah
Capacity: 100 discharges at 200 Joules or 120 discharges at 150 Joules
Shelf Life (25°C±15°C): 10 years (5 years storage + 5 years standby) 5 years standby (after installation)
Electrical Isolation
Power: Unit operates on internal battery only
External Electrical Connections:
No external devices are attached to the unit
Risk Current Category: Internally powered equipment with defibrillator-proof BF type patient applied part (as per definition of IEC 60601-1 standard)
Physical
Dimensions 300 X 210 X 80 mm
Weight 2 kg